4. Carefully separate the parts (see
Yoke
Block
5. Use tweezers to remove the old cell gasket from the cell body (see
Figure
Alignment
Figure 9-55. ED Cell Gasket for Conventional Working Electrodes
6. Rinse the surface of the cell with ASTM Type I (18 megohm-cm) filtered
and deionized water that meets the specifications listed in
7. Clean the polished surface of the cell with a clean, damp lint-free tissue.
8. Install the new gasket over the alignment pins on the cell body. When
correctly installed, one end of the gasket extends beyond the cell body, to
facilitate gasket installation and removal.
9. Verify that the gasket is flat against the cell body and is not wrinkled.