Thermo Scientific Dionex ICS-5000 Operator's Manual page 359

Ion chromatography system
Table of Contents

Advertisement

Conditioning the Capillary EGC
Capillary
IC
1. Use a high-pressure (blue) fitting bolt and ferrule to connect the
line from the
Figure
2. Disconnect the lines from the
TC.
3. Connect a piece of 0.25 mm (0.010 in) ID PEEK tubing between the
ports. This is a temporary connection.
(0.0025 in) ID
PEEK Tubing
Figure 9-21. Flow Schematic for Conditioning the Capillary EGC
4. Verify that the suppressor Mode is off.
Doc. 065342-06 12/11
ELUENT IN
port of the CR-TC to the EGC
.
9-21)
EGC
(Capillary)
0.062 mm
Assembly
EGC
(P/N 072204)
OUT
To CR-TC
Regen Out
(blue label)
To CR-TC
Eluent In
(red label)
To CR-TC
Regen In
(orange label)
ELUENT OUT
REGEN IN
and
Capillary EGC Conditioning:
Pump at 0.02 mL/min for 15 min
EGC concentration at 50 mM
EGC
IN
ELUENT
IN
To CR-TC
Eluent Out
(yellow label)
9 • EG Service
EGC OUT
OUTLET
port (see
port of the CR-
TO PUMP/DAMPER
1.58 mm (0.062 in) ID
Tefzel Tubing
Assembly (P/N 072225)
0.25 mm (0.010 in)
ID
PEEK Tubing
(Temporary
connection)
343