Dionex ICS-5000 Operator's Manual
HAUTE TENSION—Débranchez le cordon d'alimentation principal de sa
source et du panneau arrière du TC.
HOCHSPANNUNG—Ziehen Sie das Netzkabel aus der Steckdose und
der Netzbuchse auf der Rückseite des TC.
3. The fuse cartridge is located next to the main power switch (see
Figure
4. Replace the two fuses with new 2 A IEC 60127-2 time lag fuses
(P/N 954776). Thermo Fisher Scientific recommends always replacing
both fuses.
5. Reinstall the fuse cartridge.
6. Reconnect the main power cord and turn on the power.
414
9-78). Use a small screwdriver to remove the fuse cartridge.
Figure 9-78. TC Fuse Cartridge
Fuse Cartridge
Doc. 065342-06 12/11