Performance Tests

3. Receiver Minimum R Channel Level

3. Receiver Minimum R Channel Level
Perform this test to verify the minimum R channel input power level at which phase lock
can be accomplished.
Analyzer warmup time: 30 minutes.
Specifications
Frequency Range
300 kHz–3 GHz
a
3 GHz–6 GHz
a. Only for analyzers with Option 006.
Required Equipment
Description
Cable Type-N (m) to Type-N (m) 24-inch
Attenuator, 30 dB
Procedure
1. Connect the equipment as shown in
Figure 2-5
Receiver Minimum R Channel Level Test Setup
2-12
Minimum R Channel Level
−35 dBm
−30 dBm
HP/Agilent Part or Model Number
11500B
8491A Option 030
Figure
2-5.