Basic Operations for RF Devices Measurement
STEP 3. Calibration

Measuring LOAD Calibration Data

Use the 50 Ω (LOAD) standard to perform LOAD calibration by following these steps.
Step 1. As shown in Figure 3-35, turn the outside connector nut of the 50 Ω (LOAD) standard
counterclockwise to fully retract the inside connector sleeve.
Retracting connector sleeve of 50 Ω (LOAD) standard
Figure 3-35
Step 2. As shown in Figure 3-36, turn the 50 Ω (LOAD) standard clockwise with the provided
torque wrench to connect it securely to the 7-mm terminal.
Connecting 50 Ω (LOAD) standard
Figure 3-36
Step 3. Click the
Figure 3-37
Step 3
Step 4. A √ mark appears on the left side of the
calibration data measurement.
Step 5. Turn the 50 Ω (LOAD) standard counterclockwise to remove it.
64
button (Figure 3-37).
Meas Load
button upon completion of the LOAD
Meas Load
Chapter 3