Load The Default Calibration Factors - Agilent Technologies HP 1660E Series User Manual

Logic analyzers
Table of Contents

Advertisement

NOTE:
1 Go to the Scope Calibration menu.
2 Select the Mode field, then select Service Cal from the pop-up
3 Select the Procedure field, then select Default Values from the
4 Select the Start field and follow the instructions on the display.
Using the Oscilloscope
Calibrating the oscilloscope

Load the default calibration factors

Note that once the default calibration factors are loaded, all
calibrations must be done. This includes all of the calibrations in the
Self Cal menu. The calibration must be performed in the exact
sequence listed below.
The calibration PROTECT/UNPROTECT switch on the back of logic analyzer
must be set to UNPROTECT.
menu.
pop-up menu.
Loading the Default Calibration Factors
After you select the Start field, you can abort the calibration procedure
by selecting either the Mode or Procedure fields if the Continue field is
still displayed on the screen.
104

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 1660es seriesHp 1660ep series1670e series

Table of Contents