Setting The Range - Agilent Technologies E4419B User Manual

Dual channel power meter
Hide thumbs Also See for E4419B:
Table of Contents

Advertisement

Power Meter Operation

Setting the Range

Setting the Range
The power meter has no internal ranges which can be set. The only ranges
that can be set are those of the Agilent E-series power sensors. With an
Agilent E-series power sensor the range can be set either automatically or
manually. Use autoranging when you are not sure of the power level you
will be measuring. There are two manual settings either "LOWER" or
"UPPER". The lower range is more sensitive than the higher range.
"LOWER" covers the range of -70 dBm to -13.5 dBm. "UPPER" covers the
range of -14.5 dBm to +20 dBm. The default is "AUTO".
Note
To set the range:
2-58
The following procedure details how to set the range for channel
A. To set the range on channel B use the same procedure using the
B Input Settings
System
1. Press
,
A Input Settings
Inputs
range is displayed under the
2. To change this setting press
Use
or
3. To confirm your choice press
softkey.
. The current setting of the
softkey.
Range
. A pop up window appears.
Range
to highlight your choice.
.
Enter
Agilent E4419B User's Guide