Setting the Range

The power meter has no internal ranges which can be set. The only ranges that
can be set are those of the Agilent E-Series power sensors or N8480 Series
power sensors (excluding Option CFT). With an Agilent E-Series power
sensors or N8480 Series power sensor (excluding Option CFT), the range can
be set either automatically or manually. Use autoranging when you are not
sure of the power level you will be measuring. There are two manual settings
either "LOWER" or "UPPER". The lower range is more sensitive than the higher
range. "LOWER" covers the range of –70 dBm to –13.5 dBm. "UPPER" covers the
range of –14.5 dBm to +20 dBm. The default is "AUTO".
The following procedure details how to set the range for channel A. To set the range on
N O T E
channel B, use the same procedure using the
To set the range:
1 Press
2 To change this setting, press
3 To confirm your choice, press
E4419B Power Meter User's Guide
A Input Settings
,
displayed under the
to highlight your choice.
B Input Settings
. The current setting of the range is
softkey.
Range
Range
. A pop-up window appears. Use
Enter
.
Power Meter Operation
softkey.
2
or
85