HEIDENHAIN POSITIP 8000 Series Operating Instructions Manual page 199

Digital readout
Table of Contents

Advertisement

Information for OEM and Setup users | Servicing and maintenance
8.3
Maintenance plan
The product is largely maintenance-free.
Operating defective devices
Operating defective devices may result in serious consequential damage.
Do not repair or operate the device if it is damaged
Replace defective devices immediately or contact a HEIDENHAIN service
agency
The following steps are only to be performed by electrical specialists.
Further information: "Personnel qualification", Page 15
Maintenance step
All labels and symbols
provided on the product
must be checked for
readability
Inspect electrical
connections for damage
and check their function
Check power cables for
faulty insulation and
weak points
8.4
Resuming operation
When operation is resumed, e.g. when the product is reinstalled after repair or
when it is remounted, the same measures and personnel requirements apply as for
mounting and installing the product.
Further information: "Mounting", Page 63
Further information: "Installation", Page 69
When connecting the peripheral devices (e.g. encoders), the operating company
must ensure safe resumption of operation and assign authorized and appropriately
qualified personnel to the task.
Further information: "Obligations of the operating company", Page 16
8.5
Updating the firmware
The firmware is the operating system of the product. You can import new versions
of the firmware via the product's USB port or the network connection.
Prior to the firmware update, you must comply with the release notes
for the respective software version and the information they contain
regarding reverse compatibility.
In order to be on the safe the side, the current settings must be backed
up if the unit's firmware is going to be updated.
HEIDENHAIN | POSITIP 8000 | Operating Instructions Turning | 03/2022
NOTICE
Interval
Corrective action
Annually
Contact HEIDENHAIN service
agency
Annually
Replace defective cables.
Contact HEIDENHAIN service
agency if required
Annually
Replace power cables
according to the specification
II
199

Advertisement

Table of Contents
loading

Table of Contents