Saving/Recalling Instrument Setup State (Save/Recall Function); Overview Of Save/Recall Functionality - Keysight E4981A User Manual

120 hz/1 khz/1 mhz capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Setting up Measurement Conditions and Display

Saving/Recalling Instrument Setup State (Save/Recall Function)

Saving/Recalling Instrument Setup State (Save/Recall Function)
You can save/recall up to 20 instrument setup states into/from the built-in
FLASH memory (0 through 9) and external USB memory (10 through 19).
The ninth instrument setting corresponds to Auto Recall. Auto recall is not
executed when power is ON, by pressing the [Preset] key.

Overview of save/recall functionality

Configurations and measurement results can be saved into, and recalled from,
the E4981A's internal memory or external USB memory through the save/recall
functionality.
Save methods and their uses
Table 4-8
Table 4-8
Save methods and their uses
For information on configuration states that can be saved, see
"Initial settings, settings that can be saved/recalled, settings that can be
backed up," on page
76
shows available save methods and their uses:
Save method
File format
Type
(extension)
Configuration
Save (internal
-----------
memory)
Configuration
Save (USB
memory)
Data Save (USB
CSV format (.csv)
memory)
Screen Save
GIF format (.gif)
(USB memory)
C-228.
Recallable
Yes
(.sta)
Yes
No
No
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Use
To save E4981A's configuration
states into the internal memory.
To save E4981A's configuration
states into USB memory.
To save measurement results into
USB memory.
To save E4981A's screenshot into
USB memory.
Table C-1,

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4981a-001E4981a-002

Table of Contents