D
Error Messages
The Keysight E4980A/ E4980AL provides error messages to indicate its
operating status. This appendix describes the error messages of the E4980A/
E4980AL in alphabetical order.
Error Messages
An error message is displayed in the instrument status display area in the
lower-left part of the screen. Pressing the cursor arrow key on the front panel
or executing the :DISP:CCL command clears the error message. Moreover,
about a specific error message, when a sweep is started again, the display of
an error message may disappear. Errors caused by operation of a front panel
key simply appear on the display; with a few exceptions, these are not stored in
the error queue.
The log of an error message, a maximum of 100 pieces are recorded, and it can
go back and check from the oldest error. Executing the :SYST:ERR? command
checks error logs. This operation can be performed only by "SCPI" command. It
cannot be operated from the front panel.
An error with a positive error number is one uniquely defined for this
instrument. On the other hand, an error with a negative error number is
basically one defined for common GPIB devices in IEEE488.2.
A
1103
A1 EEPROM write error
An error is generated while writing data to A1 EEPROM. When this error
occurs, contact Keysight Technology's Sales and Service Office or the
company from which you bought the device.
B
62
Bias off, Turn bias on
DC bias sweep is performed with the DC BIAS function OFF.
-160
Block data error
Block data is improper.