The Keysight E4981A provides error messages to indicate its operating status.
This appendix describes the error messages of the E4981A in alphabetical
order.
Error Messages (Alphabetical Order)
Error messages are displayed in the lower row of the E4981A's display. You can
read them out by using the SCPI command. This section provides a description
of each error message and its remedy.
Errors with a negative error number are basically general errors defined by
IEEE488.2 for GPIB instruments. On the other hand, errors with a positive
error number are defined specifically for the E4981A.
1 - 100
21
1 MHz opt. not installed
This error occurs when cable correction commands are sent to E4981A with
Option 002, via GPIB/LAN/USB. This error is not generated with front panel
operation.
A
1103
A1 EEPROM write error
An error is generated while writing data to A1 EEPROM.
Contact Keysight Technology's Sales and Service Office or the company from
which you purchased the device.
B
-168
Block data not allowed
A block data element has been received where the E4981A does not accept
any block data element.