Messages

Error Messages

Error Messages
An error message is displayed against a red background in the instrument
message/warning area in the lower left part of the screen. Pushing a front
panel key or executing :DISP:CCL command clears the error message. Errors
caused by the operation of a front panel key simply appear on the display. They
are not stored in the error queue with some exceptions.
An error with a positive error number is one uniquely defined for this
instrument. On the other hand, an error with a negative error number is
basically one defined for common GPIB devices in IEEE488.2
A
20
Additional standard needed
The GPIB command that turns ON the calibration function has been sent
before all of the data measurements needed to calculate the calibration factor
have been completed. In 1-port calibration, for example, when measurements
completed for OPEN and SHORT standards but not yet for LOAD standard,
You have tried to turn on the calibration function by calculating 1-port
calibration coefficient using SENS:CORR:COLL:SAVE or
SENS:CORR:OFFS:COLL:SAVE.Be sure to measure all necessary calibration
data before sending commands.This error is not generated by front key
operations.
B
-168
Block data not allowed
An block-data element has been received at a position where this instrument
does not accept one.
C
240
Calibration data lost
This error occurs when a file containing the system calibration data is not
found or in a damaged state at time of the startup of this instrument, indicating
a failure of this instrument. Contact an Keysight Technologies sales office or
the company from which you bought the instrument.
22
Calibration method not selected
This error occurs when the command for validating the calibration data,
SENS:CORR:COLL:SAVE or SENS:CORR:OFFS:COLL:SAVE is executed before
the command for selecting a calibration type, SENS:CORR:COLL:METH:xxxx or
SENS:CORR:OFFS:COLL:METH:xxxx, is executed.This error is not generated by
front key operations.
342
Keysight E5071C Service Manual