Technology > Embedded RDS > Troubleshooting
Transmission timing
Detailed procedure
When the Web browser
1) When the power is
option license is entered /
turned ON, check the
transferred
license condition of the
WEB browser option.
(However, the case when
2) Send browser
"disabling of service browser
information to UGW
[OFF]" is set is excluded)
when the license status
is OFF=>ON due to
entry of a license or
when it is ON=>OFF
due to license transfer.
* For detailed procedures, see "Steps to Service Browser settings".
2
Technology > Embedded RDS > Troubleshooting
Transmission
Error occurs
information
Service browser mode:
Retransmission
Enabling [Active] or
is repeated
until it finishes
Stopping [Suspend]
successfully.
WEB browser
option: [ON] or [OFF]
according to the
license status
T-2-118
T-2-118
Troubleshooting
No.1
Symptom: A communication test (COM-TEST) results NG!
Cause: Initial settings or network conditions is incomplete.
Remedy1: Check and take actions mentioned below.
1) Check network connections
Is the status indicator LED for the HUB port to which the main unit is connected ON?
YES: Proceed to Step 2).
NO: Check that the network cable is properly connected.
2) Confirm loop back address
(a) Select [Settings / Registration (User Mode)] > [Preferences] > [Network] > [TCP/IP
Settings] > [IPv4 Settings] > [PING Command], enter "127.0.0.1", and touch the [Start]
button.
Is the response from the host displayed? (see following figure)
YES: Proceed to Step 3).
NO: There is a possibility that the main unit's network settings are wrong. Check the
details of the IPv4 settings once more.