LEAKING- 7.0.0 "SAMPLE BAG CONNECTED?" / ISOLATE DIALYSATE SAMPLING OPTION
A leak was occurring in a Cleaning Program (Rinse, Heat Disinfect, etc.) and you saw the "Sample Bag
Connected?" banner on the screen.
a) Turn the machine on. When "Press CONFIRM For Service Mode" appears press 'CONFIRM'. The screen
says "Machine In Service Mode".
b) Select the screen's 'Options' button.
c) Select the screen's 'Hardware Options' button.
d) Does the "Yes" box at Dialysate Sampling (Figure right) have a blue 'X'
in it?
Yes
Dialysate Sampling = "Yes"! See procedure number
No
Dialysate Sampling = "No"! This is NORMAL and is not causing the leak! Turn the machine
OFF then return to (ABOVE) procedure number

LEAKING- 7.1.0 DIALYSATE SAMPLING = "YES"

a) Select Dialysate Sampling.
b) Place the 'X' in the "No" box and press 'CONFIRM'.
c) ENSURE the "No" box at Dialysate Sampling now has a blue 'X' in it!
d) Reattach and all 'blown off' tubing but
e) Turn the machine OFF then back on!
f)
From the 'Select Program' screen select RINSE then press 'CONFIRM'. The "Sample Bag Connected"
banner should NOT appear!
g) With the machine in RINSE Program NOTE: Squirting from the Fluid Sample Connector is NORMAL =
UF Check Valve #63 is oriented correctly!
h) If (and ONLY if) an active leak reoccurs turn the machine OFF then return to (ABOVE) procedure
number
LEAKING- 1.1.0
LEAVE
the Fluid Sample Connector in the bucket!
(page 555).
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
LEAKING- 7.1.0
LEAKING- 1.1.0
(page 555).
(page 569).
Page 569