Co- 1.0.3.7 / Conductivity Low - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

Procedure # CO- 1.0.3 continued:
H) FIRMLY connect BOTH concentrate connectors to
Return to Dialysis Program ("Select Program" → 'Dialysis' → 'CONFIRM')
I)
J) From the Home screen, set [Dialysate Flow] to 800 ml/min and press 'CONFIRM'.
K) See procedure number

CO- 1.0.3.7 / CONDUCTIVITY LOW

A) Figure right, while watching through the inlet tubing,
disconnect the ACID connector from the jug, for seven (7)
seconds THEN FIRMLY reconnect!
B) Watching for thirty (30) seconds, TWO (2) possible
scenarios:
1)
IF (and ONLY if) air / acid is moving in one
direction ONLY, towards the machine!
Continue to part C.
2) IF air / acid is NOT moving OR moving back and
forth! Proceed to
CO-
4.1.0.
C) Figure right, watching through the inlet tubing,
disconnect the BICARB connector from the jug, for
seven (7) seconds THEN FIRMLY reconnect!
D) Watching for thirty (30) seconds, TWO (2) possible
scenarios:
1)
IF (and ONLY if) air / bicarb is moving, in one
direction ONLY, towards the machine! See
procedure number
2) IF air / bicarb is NOT moving OR moving back and forth! Proceed to
number
CO-
4.1.0.
CO- 1.0.3.7
(page 322).
page 345
, procedure number
page 323
CO- 1.0.3.8
(
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
JUGS
!!
).
.
page 345
, procedure
Page 322