FIL- 1.0.1 ISOLATE AIR REMOVAL SYSTEM

a) Turn the machine OFF!
b) The Deaeration gauge is used next. ENSURE it reads 0 inHg before installing it!
c) Figure below, tee the gauge into the Inlet (clear tubing) side of the Deaeration Pump!
d) Turn the machine on and return to Dialysis Program ("Select Program → "Dialysis' → 'CONFIRM).
e) Is Deaeration Pressure OKAY? Refer to
Yes
Deaeration Pressure OKAY! See procedure number
No
Deaeration Pressure in NOT OKAY! ENSURING a "No Water" alarm is NOT presenting,
NOTE this page number, as you will return here, THEN proceed to
- DEAERATION

FIL- 1.0.2 DEAERATION PRESSURE OKAY

a) Turn the machine OFF.
b) Remove the gauge and reattach the tubing.
c) Turn the machine on and return to Dialysis Program ("Select Program → "Dialysis' → 'CONFIRM').
d) See procedure number
Appendix A
PROBLEMS.
FIL- 1.0.3
(page 413).
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
(page 667) for what pressure should be.
FIL- 1.0.2
(page 412).