Fresenius Medical Care 2008 K Service Bulletin
Fresenius Medical Care 2008 K Service Bulletin

Fresenius Medical Care 2008 K Service Bulletin

Hemodialysis machines
Hide thumbs Also See for 2008 K:

Advertisement

Quick Links

Fresenius Medical Care North America
Equipment:
2008
Bulletin:
15-FRK-001 Rev A
Subject:
Optional Shunt Box Ferrite Core Installation (Update)

1.0 PURPOSE

To inform all Facilities, Field Service and Technical Support personnel of how to replace the
shunt box wire harness (which includes a ferrite core) for ESD related "Dial Valve Failure 1"
alarms on a 2008K Hemodialysis machine. This Service Bulletin replaces FSB# 08-FRK-003.

2.0 PARTS REQUIRED

3.0 REQUIRED TOOLS

#1 Phillips Screwdriver
#2 Phillips Screwdriver

4.0 INSTRUCTIONS

4.1
Stabilize the machine by applying the caster brake.
4.2
Remove any items hanging on the I.V. pole.
4.3
Locate and remove the screw attaching the I.V. pole to
its mounting base using a 5mm Hex (Allen) Wrench.
See Figure 2 for screw location.
4.4
Slide the I.V. pole up and out of the shunt box.
© 2015, Fresenius Medical Care North America, All rights reserved.
Service Bulletin
®
K Hemodialysis Machines
Figure 1 – Shunt Switch Assembly (P/N M42408)
FSB# 15-FRK-001 Rev. A
Page 1 of 7
Masking tape or equivalent
5mm Hex (Allen) Wrench
I.V. Pole Mounting
Screw Location
Figure 2

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 2008 K and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Fresenius Medical Care 2008 K

  • Page 1 See Figure 2 for screw location. Slide the I.V. pole up and out of the shunt box. I.V. Pole Mounting Screw Location Figure 2 © 2015, Fresenius Medical Care North America, All rights reserved. FSB# 15-FRK-001 Rev. A Page 1 of 7...
  • Page 2 With the machine turned off, remove the two (2) screws securing the card cage into the cabinet (see Figure 3). Slide the card cage out the front of the cabinet a few inches. Figure 3 Note: The middle panel has a ground wire attached. Take care not to damage this wire when removing the panel.
  • Page 3 Once all four screws have been removed, the shunt assembly will be free of the cabinet. However, there will be a ground wire still attached within the shunt assembly. The ground wire screw (white arrow in Figure 6) can only be removed once the shunt box is free of the cabinet.
  • Page 4 4.12 With the shunt door still closed, locate and remove the shunt switch cover screw using the arrow in Figure 9 as reference. Retain this screw. Figure 9 – Shunt Switch Cover 4.13 Carefully open the shunt door so the shunt switch cover can be removed out of the top of the shunt assembly.
  • Page 5 4.15 Once the mounting screw is removed, the shunt switch assembly can be removed out of the top of the shunt box structure. See Figure 12. Figure 12 – Switch Assembly Removal 4.16 With the shunt door still open, slide the new shunt switch assembly (P/N M42408) into the shunt box structure using the noted orientation during removal.
  • Page 6 4.18 Slide the shunt switch cover over the switch assembly and out the back of the shunt box structure as shown in Figure 15. Figure 15 – Shunt Switch Cover Protrusion 4.19 Once the cover is seated properly, close the shunt door and align the shunt switch cover over the threaded hole in the shunt switch assembly bracket.
  • Page 7 Once the machine is out of bypass, open the shunt door and confirm the machine goes into bypass by observing the float in the flow indicator. If there are any questions regarding this bulletin, contact Fresenius Medical Care Technical Support at 800-227-2572 FSB# 15-FRK-001 Rev. A...