Clean- 1.0.1 Isolate Restriction; Clean- 1.0.3 Leak Occurs - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

c)
ENSURE the 'to drain' tubing is not kinked AND (if
used) the 'Quick connector' is attached PROPERLY
to the station.
till INSTRUCTED OTHERWISE
d) Figure right,
direct the Fluid Sample Connector into a bucket on
the floor!
e) Below the machine will be returned to a Cleaning
Program. If (and ONLY if) a leak reoccurs DO NOT
turn the machine off but instead press 'Escape' then
'CONFIRM' to Interrupt the program!
f)
See procedure number

CLEAN- 1.0.1 ISOLATE RESTRICTION

a) If the problem was occurring in Heat Disinfect return to it. If not place the machine in RINSE.
b) TWO (2) checks:
CHECK #1: Figure right, water from the Fluid Sample port = UF Check
Valve #64 is bad!
CHECK #2: No 'squirts' into the bucket = UF Check Valve #63 is bad!
c) Allow up to four (4) minutes
scenarios below:
1)
IF (and ONLY if) NO leaks! Proceed to
2) IF a leak reoccurs! With the "Interrupted" banner up, see procedure number
156).

CLEAN- 1.0.3 LEAK OCCURS

a) Call debug screen 0 (Figure
right). If debug does not
appear press 'Escape' then
call screen 0.
b) Ignoring Flow Error watch
nd
2
window down
the
Valve Error
for one (1)
minute. TWO (2) possible
scenarios:
1)
IF (and ONLY if) Valve Error = 0 OR 'blinks to 1' for LESS THAN two (2) seconds: proceed
page 562
to
2)
IF (and ONLY if) Valve Error EVER = 1 MORE THAN (2) seconds: Proceed to
procedure number
CLEAN- 1.0.1
(page 156).
OR
until if a leak reoccurs. TWO (2) possible
,
, procedure number
LEAKING-
CLEAN-
7.0.0.
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
,
page 157
, procedure number
4.0.0.
CLEAN-
1.0.5.
CLEAN- 1.0.3
(page
page 189
,
Page 156