CO- 1.0.9 VERIFY DIALYSATE LIMITS

a) Select the 'Dialysate' tab (Figure below).
TCD
b)
is based on Dialysate Composition and indicates what the 'Actual' Conductivity would be if
everything was perfect.
c) Using [Alarm Width], spread Conductivity Limits as wide as they will go and press 'CONFIRM'.
d) Using [Alarm Position], set the upper Limit to 0.5 above TCD and press 'CONFIRM'.
e) Based on the 'Actual' [Conductivity] window (Figure right),
TWO (2) possible scenarios:
1)
IF (and ONLY if) [Conductivity] is STABLE between 13.4
and 14.3 mS: See procedure number
2) ALL OTHERS: Proceed to
CO-
2.0.0.
Dialysate SCREEN
CO- 1.0.10
page 337
, procedure number
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
(page 336).
Page 335