PHT- 3.1.5 ALL OTHER SCENARIOS / PRESSURE TEST FAILED
®
Has the DiaSafe
filter been replaced in THIS Troubleshooting session?
Yes
The DiaSafe
(PRESSURE) PROBLEMS
No
a) Replace the DiaSafe
Pressure Tests.
b) Place the machine into RINSE for five (5) FULL minutes!
c) Return to Dialysis Program ("Select Program" → 'Dialysis' → 'CONFIRM').
d) From the Home screen, set [Dialysate Flow] to 800 ml/min and press 'CONFIRM'!
e) Allow the machine to come out of bypass i.e. the external flow indicator's 'bob' moving up
and down in the sight glass!
f)
Allow one (1) minute BEFORE continuing to part g.
g) Reset all alarms and repeat the Pressure Tests. Do they BOTH pass?
Yes
No
®
filter has been replaced! Proceed to
®
filter preferably with one from another machine that is passing
Both Pass! The DiaSafe
HOWEVER, IMPORTANT!
484).
Pressure test(s) Fail! Proceed to
PROBLEMS
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
page 440
®
filter MAY have been causing the problem
See procedure number
page 440
,
SECTION 9 – TMP (PRESSURE)
,
SECTION 9 – TMP
PHT- 4.0.0
(page
Page 483