DAIR- 4.0.0 ISOLATE VALVE #39

If Valve #39 is sticking open target Deaeration Pressure CANNOT be achieved:
a) Per the Figures above AND right locate
Valve #39 on the BOTTOM of the
hydrochamber.
b) Figure right, clamp Valve #39's clear
tubing at the location seen.
c) Can [PUMP RATE] now be adjusted to
achieve target pressures?
Yes
Target pressures achieved! See procedure number
No
Pressures CANNOT be achieved! See procedure number
Figure 84 – Hydraulics Rear View
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
DAIR- 4.0.1
(page 501).
page 502
DAIR- 5.0.0
(
).
Page 500