T- 7.2.0 TEMPERATURE CONTROL CHECKS

a) Return NTC #3's connector to the
b) ENSURE NTC #2's connector is in the
Turn the Heater Switch ON!
c)
d) Install the hydraulics into the cabinet however, there is no need to screw the rear panels in!
e) Enter Service Mode → Options → Hardware Options. Does the "Yes" box at
'Temp Comp' have an 'X' in it?
Yes
Place the 'X' in the "No" box and press 'CONFIRM'. The 'X' turns
blue. Post a note that 'Temp Comp' is off then see procedure
number
No
Temp Comp = No! See procedure number
nd
2
distribution board position from the left, "MON-NTC".
1st
distribution board position from the left, "CON-NTC".
T- 7.2.1
(page 305).
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
T- 7.2.1
(page 305).