F- 16.0.9 MORE THAN 130 ML EVERY MINUTE / ISOLATE FLOAT SWITCH #5
a) This procedure requires up to five (5) minutes to prevent labor-intensive error!
b) Direct the tubing into the bucket for up to
130 ml every minute OR 2) Stop.
1)
IF (and ONLY if) flow continues at more than 130 ml per minute! The float is okay! See
procedure number
2)
IF flow STOPS! THREE (3) possible bad components: 1) Reoccurring debris in Restrictor #48 OR;
2) Bad Float Switch
a
To LOCATE the Float Switch refer to
five (5) minutes
F- 16.2.0
(page 125).
a
OR; 3) Bad hydrochamber
Figure 28
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
. Flow will either: 1) Continue at more than
(page 129)