Clean- 2.0.4 More Than 200 Ml Per Minute/Valve #39 Is Okay - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

Yes More than 200 ml! Valve #39 is okay! See procedure number
No Less than 200 ml! Valve #39 may be bad! See parts a AND b below:
a) ENSURE Valve #39 is plugged properly into distribution board position V39!
b) Swap in the following components (see Component List below) one at a time, with known
good, then repeat (ABOVE) procedure number CLEAN- 2.0.3 (page 167). Whhen more
than 200 ml is collected the last component swapped in is the problem.
COMPONENT LIST (4 components): 1) Actuator-Test Board; 2) Valve #39 (located on
the bottom of the hydrochamber, refer to the Figure above), including its blue wiring
harness; 3) ACTUATOR board ribbon cable; 4) Distribution board.
CLEAN- 2.0.4 MORE THAN 200 ML PER MINUTE/ VALVE #39 IS OKAY
If all procedures were performed correctly you have determined good flow through Restrictor #48 AND
Valve #39 i.e. more than 300 ml per minute to the Deaeration Pump. See parts a and b below:
Remove the clamp from Restrictor #48's tubing!
a)
page 165
b) Return to
, procedure number
CLEAN- 1.6.0.
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
CLEAN- 2.0.4
(page 168).
Page 168