Scenario #3: Referring to Figure C, for a ONE MINUTE PERIOD if EVER a
continuous weak stream that LASTS for five (5) seconds or
longer:
58
Figure C
Scenario #4: IF No Flow EVER:
F- 6.1.0 PULSE – STOP – PULSE OR 'DRIBBLE' / ISOLATE FOR STUCK OPEN BC VALVE
A) Reattach the drain tubing to the station!
B) ENSURING Dialysate Flow in ON (Flow on/off lamp is NOT blinking), ENSURE a good drain
connection by listening for or visually verifying flow.
C) If a TMP alarm is NOT present skip to part D. If a TMP alarm is present: 1) Press and release the
'Reset' key then immediately press and hold it for three (3) seconds; 2) Allow thirty (30) seconds;
3) If a TMP alarm reoccurs attempt RESET up to twice more BEFORE continuing to part D.
D) Allow thirty (30) seconds BEFORE continuing to part E!
E) Call debug screen 4 to watch PDIA (left column) AND ADIA (right
column) for TWO (2) minutes. BOTH should REMAIN between
1.5 and 8.5 AND STABLE i.e. DOES NOT change more than 0.6
per minute!
F) TWO (2) possible scenarios:
1)
IF (and ONLY if) PDIA AND ADIA is STABLE between 1.5 and 8.5: Proceed to
procedure number
2) IF PDIA and / or ADIA
procedure number
ENSURING the drain tube OPENING was pointed up, proceed to
, procedure number
F-
6.2.0.
If (and ONLY if) the Dialysate Flow on/off lamp is blinking
return to (ABOVE) procedure number
NOT blinking proceed to
F-
6.3.0.
F-
6.1.4.
IS NOT
remaining between 1.5 and 8.5 and / or
F- 6.1.1
(page 52).
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
page 49
F- 5.2.0
(
page 59
, procedure number
page 54
IS NOT
stable: See
page
). If
,
Page 51

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2008k

Table of Contents