T- 1.4.3 Temperature Less Than 30 Degrees C/Calibrate Temp Control/Confirm Sulution - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

Yes
Less than 30.0° C! See procedure number
No
More than 30.0° C! Swap the following components (see COMPONENT LIST below) in one at
a time, with known good, in between repeating (ABOVE) procedure number
244) until Temperature remains less than 30.0° C.
COMPONENT LIST: 1) Actuator-Test Board
4) Functional Board (possibly IC20)
1
To LOCATE these boards refer to
2
To prevent "Cond Offset Failure" place the machine into T and C Mode (refer to
OPERATING MODES
T- 1.4.3 TEMPERATURE LESS THAN 30.0° C / CALIBRATE TEMP CONTROL / CONFIRM SOLUTION
a) Return NTC #3's connector to the
position from the left, "MON-NTC".
b) Perform BOTH procedures below:
Procedure #1: Calibrate Temp Control per the
Calibration Procedures
booklet!
Procedure #2: When the calibration is complete continue to part c.
c) Turn the machine OFF!
d) Turn the machine on and return to Dialysis Program ("Select Program" → 'Dialysis' → 'CONFIRM')!
e) Again unplug NTC #3 from the
f)
Call the Home screen. Based on the [Temperature] window, TWO (2) possible scenarios:
1)
IF now less than 30.0° C: Return NTC #3's connector to distribution board position "X3, MON-
NTC" then allow ten (10) full minutes. If a temperature problem reoccurs return to (ABOVE)
procedure number
problem is solved!
2) IF more than 30.0° C: Swap the following components (see COMPONENT LIST below) in one at a
time, with known good, and in between repeat procedure number
[Temperature] window is less than 30.0° C.
COMPONENT LIST: 1) Actuator-Test Board
4) Functional Board (possibly IC20)
a
To LOCATE these boards refer to
b
To prevent "Cond Offset Failure" place the machine into T and C Mode (refer to
MODES
(page Error! Bookmark not defined.)).
1,2
; 5) Distribution board; 6) Motherboard.
Figure 4A
(page Error! Bookmark not defined.))
nd
2
distribution board
nd
2
distribution board position from the left, "X3, MON-NTC"..
T- 1.0.0
(page 215). If not perform the alarms test. If the tests pass the
a,b
; 5) Distribution board; 6) Motherboard.
Figure 4A
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
T- 1.4.3
(page 245).
1
1,2
; 2) Sensor Board
(page 9)
T- 1.4.3
a
; 2) Sensor Board cable; 3) Sensor Board
(page 9).
T- 1.4.2
(page
; 3) Sensor Board cable;
(page 245) until the
a,b
;
OPERATING
Page 245