Fresenius Medical Care 2008K2 Troubleshooting Manual page 598

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

7)
IF the machine NEVER turns on i.e. fan NEVER runs: Perform parts a THROUGH d below:
a) If the fan IS NOT running continue to part b. If the fan IS RUNNING but, after fifty (50) seconds, the
screen REMAINS BLACK proceed to
b) Press the Power button for two (2) seconds! If the fan IS NOT running perform parts
c and d. If the fan is running but, after fifty (50) seconds, the screen REMAINS
BLACK proceed to
c) Plug into a KNOWN GOOD GFI power outlet.
d) Press the Power button for two (2) seconds! If the machine still DOES NOT turn on (fan NEVER
runs), ENSURING the GFI did not 'blow', proceed to
8)
IF after being on for fifty (50) seconds the screen REMAINS fully OR partially white and possibly
displays nothing: Proceed to
9)
IF "Failed Sending Data To Actuator Board": Proceed to
10) IF after being on (fan running) for fifty (50) seconds the screen displays 'weird' data i.e.
scrambled, weird colors, abnormal lines etc.: Proceed to
11) IF after being on (fan running) for fifty (50) seconds the screen still says "Loading
Software....Please Wait": See parts a through c below:
a) To avoid pulling cables loose, GENTLY open the card cage.
b) Behind the card cage (Figure below), ENSURE the 24V POWER harness is plugged in properly.
page 599
c) Proceed to
page 598
page 598
, procedure number
page 598
, procedure number
, procedure number
P-
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
, procedure number
P-
C.0.0.
page 595
, procedure number
P-
C.0.0.
page 597
, procedure number
page 598
, procedure number
E.0.0.
P-
C.0.0.
P-
A.0.0.
P-
B.0.0.
P-
C.0.0.