CO- 6.0.0 DISTRIBUTION BOARD CHECKS / ON LINE CLEARANCE (OLC) ENABLED?
a) Figure right, inside the distribution board, TWO
checks:
CHECK #1: ENSURE position #4, "PH-PR" is
VACANT!
CHECK #2: ENSURE Conductivity Cell #7 AND
NTC #3's connectors are plugged in
properly.
b) If troubleshooting an Online Clearance (OLC) problem see part c. If NOT, proceed to
procedure number
CO-
c) Call debug screen 0.
d) Figure right, do the Post Dialyzer Cond and Temp symbols
appear (Yes or No)?
Yes
The OLC option is enabled! See procedure
number
CO- 6.1.0
The OLC option is off! A) Enter Service mode → Options → Hardware Options; B) Set OLC to
No
"Yes" and press 'CONFIRM'; C) Return to debug screen 0 to ENSURE the Post Dialyzer
Cond and Temp symbols now appear; D) Proceed to
CO-
6.2.0.

CO- 6.1.0 OLC OPTION ENABLED

a) Figure right, ENSURE Conductivity Cell #13's
AND NTC #44's connectors are plugged in
properly.
b) Open the shunt door.
c) Allow thirty (30) seconds.
d) Is screen 0's Post Dialyzer Temp more than
34.00° C?
Yes
Post Dialyzer Temp is more than 34.00° C! See procedure number
No
Post Dialyzer Temp LESS THAN 34.00° C! Proceed to
CO- 10.0.20
6.2.0.
(page 353).
.
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
page 354
page 354
, procedure number
CO- 6.2.0
page 380
, procedure number
,
(page 354).