PDIA OR ADIA is
b) Continuous flow, more than 0.2 ml per minute, from the connector?

TMP- 1.0.2 PDIA AND ADIA STABLE BETWEEN 3.0 AND 6.0

IMPORTANT!
a)
At the bottom of the screen, select the 'Dialysate' tab.
b) Press 'CONFIRM'.
c) Figure right, if (and ONLY if) the screen's 'status' bar
is NOT BLANK see procedure number
(page 444). If BLANK repeat parts a and b!
IS NOT
Yes Continuous flow! A) Return the connector to the shunt and close the door;
B) Turn the machine OFF; C) Proceed to
DIAGNOSTIC VALVE LEAK
No continuous flow! See parts a THROUGH c below:
Return the connector to the shunt AND close door!
a)
b) Call debug screen 0 to ENSURE the TOP balancing chamber valves
are REMAINING BLUE!
c) Watch Valve #41's 'dot' (Figure right) for one (1)
minute. It should stay white! TWO (2) possible
scenarios:
1)
IF (and ONLY if) Valve #41's 'dot' REMAINS white!
Proceed to
2) IF Valve #41's 'dot' cycles between white and blue: A) Turn
the machine OFF! B) Proceed to
DIAGNOSTIC VALVE LEAK
TMP- 1.0.4
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
between 3.0 and 6.0 continued:
page 527
TESTS.
page 462
, procedure number
TESTS.
,
SECTION 18A –
.
TMP- 3.0.0
page 527
,
SECTION 18A –
Page 443