Fresenius Medical Care 2008K2 Troubleshooting Manual page 328

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

Scenario #1:
Referring to Figure A (below), approximately 30 ml pulses that stop completely
between each and every cycle i.e. Pulse → Stop → Pulse → Stop: See parts a
AND b below:
a) Reconnect the drain.
b) ENSURING [Dialysate Flow] is ON and set to 800 ml/min, see procedure
Figure A
Referring to Figure B (below), approximately 30 ml Pulse → Stop → Pulse →
Scenario #2:
'Dribble' (a noticeably weaker stream that last about two (2) seconds) → etc.
Perform parts a and b below:
a) A procedure in another section is performed next. NOTE this page number as
b) BEFORE continuing to part c proceed to page
c) If a leaking Balancing Chamber diaphragm was not located in part b, return to
d) Allow four (4) minutes then see procedure number
.
Figure B
number
CO- 1.0.4
(page 325).
you may be prompted to return here!
TESTING FOR A LEAKING BALANCING CHAMBER
Dialysis Program ("Select Program" → 'Dialysis' → 'CONFIRM')!.
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
535,
to perform
SECTION 19 -
DIAPHRAGM.
CO- 1.0.4
(page 325).
Page 324