F- 2.0.10 VALVE #24'S 'DOT' IS BLUE

a) Call the Home screen to ENSURE [Dialysate Flow] has REMAINED set to 800 ml/min!!
b) Does a leak develop? Refer to
Table 1
Does a leak develop?
Yes (a leak
develops)
No leaks
F- 2.0.11 VALVE #24'S 'DOT IS BLUE BUT 'BOB' IS NOT MOVING
Call debug screen 1 to see VERR (right column, bottom). TWO (2) possible scenarios:
1)
IF (and ONLY if) VERR = 1 or more: Proceed to
ERRORS IN DIALYSIS
2)
IF VERR = 0: Proceed to
Table 2
below:
See parts a THROUGH c below:
a) Turn the machine OFF!
b) If tubing has blown off tie wrap these segments to where they
belong!
c) Proceed to
TWO (2) possible scenarios based on if the external flow
indicator's 'bob' is rising, at least ¼ way up, in the sight tube.
1)
IF (and ONLY if) rising at least ¼ way: Proceed to
40
, procedure number
2) IF NOT rising at least ¼ way: See procedure number
F- 2.0.11
PROGRAM.