EOS- 4.0.0 PUMP SYMBOL REMAINS PINK / ISOLATE THE PUMP CIRCUITS
IMPORTANT! Turn the machine OFF to prevent damage!
a)
b) Per
Figure 63
(page 398), return both pump cables to their distribution board positions.
c) See EOS- 4.0.0, STEP #1 below:

EOS- 4.0.0, STEP #1 Isolate Actuator-Test Board

IMPORTANT! Turn the machine OFF to prevent damage!
a)
b) Swap in a known good Actuator-Test Board. To LOCATE the board refer to
c) If the problem was occurring in a Cleaning Program return to it. If not return to Dialysis Program.
d) Call debug screen 0. If debug does not appear press 'Escape' then call screen 0.
e) Allow thirty (30) seconds THEN, WITHOUT LOOKING AWAY, watch for pump symbol for up two (2)
minutes. Does it remain white now?
Yes
Remains white! The previous Actuator-Test Board is bad.
No
Turns pink! Continue to EOS- 4.0.0, STEP #2 (the previous Actuator-Test Board is okay).

EOS- 4.0.0, STEP #2 Isolate the Sensor Board cable:

IMPORTANT! Turn the machine OFF to prevent damage
a)
1
b) Swap in (or check
) a known good Sensor Board cable. To LOCATE the cable refer to
9).
1
To check the Sensor Board cable, which pump is issuing the EOS error?
IF ACID:
NOTE that one (1) ACID EOS connection (ONLY) will be checked and proceed to
524
,
good continue to procedure number EOS- 4.0.0, step #3.
IF BICARB:
NOTE that one (1) BICARB EOS connection (ONLY) will be checked and proceed to
page 524
checks good continue to procedure number EOS- 4.0.0, step #3.
c) If the problem was occurring in a Cleaning Program return to it. If not return to Dialysis Program.
d) Call debug screen 0. If debug does not appear press 'Escape' then call screen 0.
e) Allow thirty (30) seconds THEN, WITHOUT LOOKING AWAY, watch for pump symbol for up two (2)
minutes. Does it remain white now??
Yes
Remains white! The previous Sensor Board cable is bad
No
Turns pink! Continue to EOS- 4.0.0, STEP #3 next page.
SECTION 17- CHECKING THE SENSOR BOARD
,
SECTION 17 - CHECKING THE SENSOR BOARD
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Figure 4A
(page 9).
Figure 4A
CABLE. If the cable checks
CABLE. If the cable
(page
page
Page 400