F- 11.0.5 Tmp Alarm?; F- 11.0.6 Isolate For A Leaking Balancing Chamber Valve - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

Yes
PDial = 0 +/- 50! A) Return the dialyzer connectors to the shunt and close the door!
B) See procedure number F- 11.0.5 (page 105).
No
PDial is NOT = 0 +/- 50! A)
blinking); B) Proceed to

F- 11.0.5 TMP ALARM?

a) Turn Dialysate Flow ON (Dialysate Flow on/off lamp STOPS blinking).
b) Call the Home screen.
c) Set [Dialysate Flow] to
d) Is the TMP window RED (TMP alarm)?
Yes TMP window IS RED! RESET SEQUENCE: A) Press and release the 'Reset' key then
immediately press and hold it for three (3) seconds. B) Allow thirty (30) seconds. C) If a TMP
alarm reoccurs perform the RESET SEQUENCE up to twice more BEFORE continuing to
procedure number
No TMP window is white! See procedure number

F- 11.0.6 ISOLATE FOR A LEAKING BALANCING CHAMBER VALVE

TMP is STABLE if (and ONLY if) the TMP window REMAINS white AND TMP does NOT change more than
60 mmHg in three (3) minutes. TWO (2) possible scenarios:
1)
IF (and ONLY if) TMP is STABLE: See procedure number
2)
IF TMP is UNSTABLE: A procedure in a different Section is performed next! NOTE this page and
procedure number (F- 11.0.6) as you may be because you may be prompted to return here. See part A
below:
A) Before continuing to part B, proceed to
VALVE LEAK
TESTS.
B) If a balancing chamber valve leak was not located in part A turn the machine off then see
procedure number
IMPORTANT!
page 107,
500 ml/min
and press 'CONFIRM'!
F- 11.0.6
(page 105).
page 527
F- 11.0.7
(page 106).
F1106
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Turn Dialysate flow ON (Flow on/off lamp NOT
procedure number F- 12.0.0.
F- 11.0.6
(page 105).
F- 11.0.7
to perform
SECTION 18A – DIAGNOSTIC
(page 106).
Page 105