Fresenius Medical Care 2008K2 Troubleshooting Manual page 237

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

1)
IF (and ONLY if) falls to less than 35.1 °C: See procedure number
T- 1.0.22
(page 234).
2) IF remains more than 35.1° BUT NEVER more than 38.8° C: Proceed to
298
, procedure number
3)
IF EVER more than 38.8° C even if only once: Perform parts a THROUGH d
below:
a) If a 6.04 KΩ plug was previously placed in distribution board position, "MON-
NTC" remove it then, avoiding the VACANT position on the right ("PH'PR")
return NTC #3's connector to the
b) If a 274 Ω plug was previously placed in distribution board position, "X7,
COND", remove it and return the Cond Cell's connector.
Using a calibrated temperature meter ENSURE the
c)
incoming water (from the RO) is less than 35.0º C
d) If (and ONLY if) incoming water is less than 35.0° C proceed to
procedure number
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
T-
6.0.0.
nd
2
position from the left ("MON-NTC").
T-
1.3.40.
page
.
page 241
,
Page 233