Clean- 6.1.0 'Air' Moving Through Valve #29 - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

Yes
Air moving ALL THE WAY through Valve #29! See procedure number
188).
No Air is NOT moving through Valve #29! See parts a AND b below:
Remove the clamp from Valve #30!
a)
b) The Troubleshooting Guide cannot locate an immediate problem. Return to Dialysis
Program and troubleshoot potential hydraulic problems from there!

CLEAN- 6.1.0 'AIR' MOVING THROUGH VALVE #29

Turn the machine OFF and LEAVE IT OFF!
a)
b) Screw a 60 ml syringe, COMPLETELY FILLED WITH AIR,
onto the Fluid Sample Port.
c) Push hard on the syringe plunger while watching Valve
#29's OUTPUT tubing. Do you see air moving through
Valve #29?
Yes
Air moving through Valve #29! Valve #29 is
bad (sticking open).
No
Air is NOT moving through Valve #29! The Actuator-Test Board is bad!
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
CLEAN- 6.1.0