CO- 10.0.405 VERIFY SENSOR VERSUS ACTUATOR

Call debug screen 5. TWO (2) checks! Proceed according to the Yes – No choice at check #2.
Check #1
CPRE (middle column) versus ACOND (upper left)): They
change continually but MUST REMAIN within +/- 500 of each
other?
Check #2
CPRE versus CPOS (middle column)): Change continually but
MUST REMAIN within +/- 500 of each other?
Yes (to both): CLOSE THE SHUNT DOOR then see procedure number
(page 386).
No (to either OR both):
page 354
Proceed to
CELLS but if (and ONLY if) you return here see procedure number
CO- 10.0.41
(page 386).
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
, procedure number
CO- 6.2.0
CO- 10.0.41
to calibrate COND
Page 385