VEC- 2.2.0 VALVE #39 IS OPEN

A) Turn the machine OFF and remove the gauge!
B) Turn the machine on and allow the "Select Program" screen to appear. What color
is the 'Dialysis' button (Gray or Blue)?
1) IF GRAY: Place the machine in RINSE. Most Valve Errors will not stop Program timing for very
long. Allow RINSE to complete i.e. [Remaining Time] = 0:00, then see part C.
2) IF BLUE: See part C.
C) Return to Dialysis Program ("Select Program" → 'Dialysis' → 'CONFIRM').
D) From the Home screen, set [Dialysate Flow] to 800 ml/min
E) Allow five (5) minutes BEFORE continuing to part F.
F) Call debug screen 1 to see VERR (right column, bottom). TWO (2) possible scenarios
1)
IF (and ONLY if) VERR = 0: For some reason the Valve Error is not occurring at this time.
2)
IF VERR = 1 OR more: See parts a AND b below:
a) Call debug screen 0.
b) Ignoring the TOP Flow Error window, WITHOUT LOOKING
AWAY, watch the
(1) minute. Ignoring a 'blink to 1', that lasts less than one (1)
second, does for
THAN two (2) seconds?
Yes
No Valve Error = 1 LONGER THAN two (2) seconds! Proceed to
26..
nd
2
window down
Valve Error
EVER = 1 for LONGER
Valve Error = 0 OR 'blinks' to 1 for less than one second! Proceed to
TROUBLESHOOTING VALVE ERRORS IN DIALYSIS PROGRAM
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H