Fil- 5.0.0 Air Seen At Chamber #69 - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

FIL- 5.0.0 AIR SEEN AT CHAMBER #69

Call debug screen 1. Look at FILACT (middle column). TWO (2) possible scenarios:
1)
IF (and ONLY if) FILACT = 0: See procedure number
2) IF FILACT = 1! See parts a THROUGH c below:
Figure below
a)
Connector's distribution board position.
b) If the plug is placed correctly FILACT eventually = 0 ALWAYS!
c) See procedure number
, place one of the resistor plugs, from the FOUR-RESISTOR SET into Air Sensor
FIL- 5.1.0
(page 433).
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
FIL- 5.1.0
(page 433).
Page 432