Leaking 5.2.0 Valve #25 Okay/Isolate Valve #26 - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

1)
IF (and ONLY if), during this procedure, tubing 'blew off' the DiaSafe
top side Balancing Chamber Valves (see Figure below): A) Turn the machine off;
B) Reattach and tie wrap all tubing; c) Repeat (ABOVE) procedure number
LEAKING- 5.1.0
2) IF tubing DID NOT 'blow off': A) Reconnect the drain tubing; B) NOTE
#25
will be checked and proceed to
LEAKING 5.2.0 VALVE #25 OKAY / ISOLATE VALVE #26
a) Press 'Escape' then 'CONFIRM' twice to call the "Select Program" screen!
b) Remove the red dialyzer connector from the shunt door and place it on the floor!
c) Close the door!
d) Per the Figure above, clamp and remove Valve #26's
e) Fill the 60 ml syringe with water then attach it to Valve #26's INPUT nozzle.
f)
Push AS HARD AS YOU CAN on the syringe plunger. Can you push ALL 60 ml through Valve #26 within
FIVE (5) seconds?
Yes
Valve #26 is okay! Return the dialyzer connector to the shunt then see procedure number
LEAKING- 5.3.0
No Significant resistance felt! NOTE
page 192
(page 564).
(page 566).
ONLY
,
TROUBLESHOOTING A
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
page 192
,
TROUBLESHOOTING A
INPUT tubing
!
VALVE #26
will be checked and proceed to
VALVE.
®
filter or the
ONLY
VALVE
VALVE.
Page 565