BC Switch goes to 897 or more, even if only once! Read before performing! Return to
page 26
, procedure number
THREE (3) possible bad components (see COMPONENT LIST below):
COMPONENT LIST: 1) Actuator-Test Board
intermittent restricted balancing chamber valves #32, #34, #36 and #38
1
To LOCATE the Actuator-Test board refer to
2
To LOCATE these valves refer to
No BC Switch IS NEVER, EVER 897 or more. A Flow Error is not presenting at this time! It may be
extremely intermittent. Return to
F- 1.0.5
but, if you return here, because of no drain flow,
page 44
, procedure number
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
1
OR; 2) Actuator cable OR; 3) Multiple
Figure 4A
(page 9);
Figure 20
(page 84)
F-
3.8.0.
2
.