Fil- 4.0.0 Filact = 0/Final 'Filling Program' Checks; Fil- 4.0.2 Isolate Spent Side Balancing Chamber Leak - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

FIL- 4.0.0 FILACT = 0 / FINAL 'FILLING PROGRAM' CHECKS
Call the Home screen. Figure right, if the TMP window is RED a TMP alarm
is present. TWO (2) possible scenarios:
1)
IF (and ONLY if) the TMP window is WHITE: See procedure number
2)
IF the TMP window is RED: Perform parts a THROUGH d below:
a) Press and release the 'Reset' key then immediately press and hold it for three (3) seconds.
IMPORTANT!
b)
c) If a TMP alarm reoccurs perform parts a and b up to twice more BEFORE continuing to part d.
d) Allow thirty (30) seconds THEN see procedure number

FIL- 4.0.2 ISOLATE SPENT SIDE BALANCING CHAMBER LEAK

TMP is STABLE if the TMP window REMAINS white AND does NOT change more than 60 mmHg in three
(3) minutes. TWO (2) possible scenarios:
1)
IF (and ONLY if) TMP is STABLE: ASSUMING all procedures were performed correctly (i.e. no
air seen at Chamber #69) AND if (and ONLY if) FILACT continues to = 1 intermittently, see
(ABOVE) procedure number
2) IF TMP is UNSTABLE: See parts a THROUGH c below:
a) A procedure, in different Section, is performed next. NOTE this page and procedure number
(FIL- 4.0.2) as you may prompted to return here. Perform parts a and b below
b) Before continuing to part b, proceed to
VALVE LEAK
c) If balancing chamber valve leak was NOT located in part a, proceed to
SECTION 9- TMP (PRESSURE) PROBLEMS
Allow thirty (30) seconds!
FIL- 3.7.0
(page 430).
TESTS.
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
FIL- 4.0.2
page 527
, to perform
FIL- 4.0.2
(page 431).
(page 431).
SECTION 18A – DIAGNOSTIC
page 440
,
Page 431

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2008k

Table of Contents