L) Call the Home screen. WITHOUT LOOKING AWAY, watch [Temperature]
AND [Conductivity] for one (1) minute. Based first on
TWO (2) possible scenarios below
1)
IF between 35.5 and 38.5° C AND STABLE i.e. does NOT changes
more than +/- 0.2 °C: Going forward, if [Temperature] becomes
unstable address it FIRST! For now see part M.
2) ALL OTHERS: Proceed to
[Conductivity]
M) Based on
1)
IF (and ONLY if) between 13.0 and 14.3 mS AND STABLE i.e. does NOT change more than
+/- 0.2: See procedure number
2)
ALL OTHERS: See part N.
Firmly connect to JUGS* of known good** bicarb AND acid!
N)
* To isolate between a Central Concentrate Delivery System (SDS) and the 2008 K/K
JUGS!
** Tested good per clinic procedures!
O) ENSURE [Dialysate Flow] is ON (Dialysate Flow on/off lamp is NOT blinking)!
R) If the acid AND / OR bicarb source was changed above allow five (5) minutes BEFORE continuing to
part S.
S) See procedure number
page 214
,
SECTION 4 – TEMPERATURE PROBLEMS
, TWO (2) possible scenarios:
CO- 1.0.0
CO- 1.0.0
(page 319).
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
[Temperature]
(page 319).
,
2
connect to
Page 318