F-1.0.2.11 Chamber #69 Says "Air" Always; F-1.0.3 Chamber #69 Says "No Air"/Troubleshooting Filling Program Failure - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

1) IF says "Air" always: See procedure number
2) IF says "No Air" OR toggles between "Air" and "No Air": See parts a AND b below:
a) Inside the distribution board, ENSURE the Air
Sensor's female connector is UNPLUGGED
th
5
position
from
b) TWO (2) possible scenarios:
1)
IF (and ONLY if) Chamber #69 now says
"Air" always: See procedure number
F- 1.0.2.11
2)
IF Chamber #69 continues to say "No Air" or toggles between "Air" and "No Air":
Proceed to
F- 1.0.2.11 CHAMBER #69 SAYS "AIR" ALWAYS
a) Call debug screen 1.
b) Allow two (2) minutes OR until if FILACT (middle column) = 1?
Yes
FILACT = 1! Proceed to
No
AFTER two (2) minutes FILACT REMAINS = 0! Look at VERR (lower right).
TWO (2) possible scenarios:
1)
IF (and ONLY if) VERR = 0: See procedure number
2)
IF VERR = 1 or more: Proceed to
F- 1.0.3 CHAMBER #69 SAYS "NO AIR" / TROUBLESHOOT FILLING PROGRAM FAILURE
Turn the machine OFF!
a)
Parts b through f next page
FROM THE LEFT!
(page 22).
page 22
, procedure number
page 24
, procedure number
page 30
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
F- 1.0.2.11
(page 22).
F-
1.0.3.
F-
1.0.4.
F- 1.0.3
, procedure number
(page 22).
F-
2.0.0.
Page 22