T- 7.6.6 ADJUST 'SLOW FLOW' MONITOR REFERENCE

a) 'Sharply' press 'CONFIRM' to turn [Monitor Reference] pale yellow / white!
b) Adjust [Monitor Reference] to match the meter's maximum (i.e. between 36.9 and 37.1° C).
c) CAREFUL HERE! Sharply 'CONFIRM' TWICE to turn [Monitor Reference] GRAY!
d) Does the GRAY [Monitor Reference] data box read between 36.4 and 37.6° C?
Yes
The GRAY data box is between 36.4 and 37.6° C! This completes the calibration HOWEVER,
if you started out troubleshooting a "TEMP OVER 95 DEGREES" alarm return to procedure
number
T- 1.5.1
No
The GRAY data box IS NOT between 36.4 and 37.8! Swap in the listed components (see
COMPONENT LIST below), one at a time, with known good, in between returning to (ABOVE)
procedure number
COMPONENT LIST: 1) NTC #3
1
To LOCATE NTC #3 refer to
page 247
(
). If not, return to Dialysis Program.
T- 7.2.1
(page 305) to test each new component.
1
; 2) Functional Board.
Figure 42
(page 218)
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 313