Troubleshooting A Valve; Tsv- 1.0.0 Check Distribution Board Connections - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

To avoid damage, turn the machine OFF!
A)
There is NO NEED to check all 17 valves! Check ONLY the valve as
B)
NOTED in the procedure that brought you here!!
C) See procedure number TSV- 1.0.0 below:

TSV- 1.0.0 CHECK DISTRIBUTION BOARD CONNECTIONS

Per the
Figure
below
a)
b) TWO (2) checks to ENSURE:
Check #1:
The ACTUATOR CABLE is plugged in SECURELY AND is not damaged!
Check #2:
The NOTED valve is plugged PROPERLY into its position. If not, this most likely is the
problem!
c) Before continuing to procedure number
Figure 36
(page 193) to LOCATE the

TROUBLESHOOTING A VALVE

, remove the distribution board cover.
Figure 34 – Distribution Board
TSV- 2.0.0
NOTED
valve.
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
(page 194), refer to
Figure 35
(page 193) OR
Page 192