Fresenius Medical Care 2008K2 Troubleshooting Manual page 462

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

Yes
-250 mmHg holds! A)
segments then; B) Leaving [Dialysate Flow] off, proceed to
TMP-
4.0.0.
No
-250 could not be achieved or did NOT hold! Either the transducer protector is wet OR a Four-
Way assembly tubing connection is leaking! See parts A and B below:
A)
Locate and repair the leak. Consider using another Four-Way assembly!
B)
Return to (ABOVE) procedure number
IMPORTANT!
Remove the clamps from both Four-Way tubing
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
page 466
TMP- 2.0.16
(page 456).
, procedure number
Page 458