BL- 2.0.0 BLOOD LEAK ALARM PRESENT (CHECK FOR AIR)

IMPORTANT!
Air in dialysate cause false blood leak alarms. Per the Figure below,
watch for air bubbles through the Blood Leak Sensor's TO AND FROM tubing for one (1) minute each?
Yes
Air seen! See procedure number
No air seen! See procedure number
BL- 3.0.0
page 516
BL- 4.0.0
(
Figure 89 – Hydraulics Side View
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
(page 514).
).
using a flashlight
,