Allow fifteen (15) seconds then watch the tubing opening for up to five (5) minutes. Is
there flow or dripping (possibly slowly) from it?
Yes
Flow or dripping seen! See procedure number
No No flow OR dripping seen! See procedure number

OVER- 4.0.1 FLOW OR DRIPPING

a) Figure right, unplug Valve #41 from
distribution board position labeled
"V27, IN-V".
b) Does flow or dripping continue from the Air
Gap tubing?
Yes
Flow or dripping continues! Valve #41* is leaking and must be replaced! *To LOCATE Valve
#41 refer to
No
Flow or dripping stops! Perform parts a
THROUGH c below:
a) Figure right, unplug the Float's connector from
distribution board position "X5, FLOAT-SW".
b) Check inside the vacant "X5" position for
corrosion or damage. Damage indicates
the distribution board needs to be
replaced!
c) Watch for up to five (5) minute. Does flow or dripping continue?
Yes
No
Figure 35
(page 193).
Flow or dripping continues! The Actuator-Test Board is bad. To located the
board refer to
Figure 4A
Flow or dripping stops! Float Switch #5 is bad! To LOCATE Float Switch #5 refer
to
Figure 28
(page 129).