CLEAN- 1.5.1 ISOLATE POSSIBLE AIR LOCKk
CAUTION!
a)
Fluid and tubing may be HOT!
b) Figure right, remove the CLEAR TUBING from the
Deaeration Pump's Input Nozzle. To encourage gravity
flow, keep the tubing at the level of the pump head!.
c) The tubing is from hydrochamber D. If any flow from it
direct it into the bucket for TWO (2) MINUTES BEFORE
continuing to part d.
d) Now direct the tubing into the cylinder for ONE (1)
MINUTE! More than three hundred (300) ml collected?
Yes
More than 300 ml! See procedure number
No Less than 300 ml! There may be an airlock between hydrochamber D and the pump. Proceed
page 166
to
, procedure number
CLEAN-
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
CLEAN- 1.6.0
(page 165).
2.0.0.
Page 164